	<item>
		<title>By: Dan</title>
		<link>http://www.potencyassay.com/2009/11/09/detecting-edge-effects/comment-page-1/#comment-11</link>
		<dc:creator>Dan</dc:creator>
		<pubDate>Thu, 29 Apr 2010 00:02:46 +0000</pubDate>
		<guid isPermaLink="false">http://www.potencyassay.com/?p=45#comment-11</guid>
		<description>Hi Yi,

The software I used to create the heatmaps was SigmaPlot, but any plotting software that can do XYZ pairs should work.

No, I don&#039;t think adding extra liquid in between the wells will do much.  If evaporation is truly a problem (I have never encountered an assay where that is true unless you incubate for a really long time), then you can use a humidity chamber to help somewhat.  There are also gas permeable membranes that you can cover your plate with.  They don&#039;t allow water vapor to pass through, but allows other gases to penetrate.

I hope that helps,
Dan</description>
		<content:encoded><![CDATA[<p>Hi Yi,</p>
<p>The software I used to create the heatmaps was SigmaPlot, but any plotting software that can do XYZ pairs should work.</p>
<p>No, I don&#8217;t think adding extra liquid in between the wells will do much.  If evaporation is truly a problem (I have never encountered an assay where that is true unless you incubate for a really long time), then you can use a humidity chamber to help somewhat.  There are also gas permeable membranes that you can cover your plate with.  They don&#8217;t allow water vapor to pass through, but allows other gases to penetrate.</p>
<p>I hope that helps,<br />
Dan</p>
]]></content:encoded>
	</item>
	<item>
		<title>By: Yi</title>
		<link>http://www.potencyassay.com/2009/11/09/detecting-edge-effects/comment-page-1/#comment-9</link>
		<dc:creator>Yi</dc:creator>
		<pubDate>Sat, 17 Apr 2010 02:22:37 +0000</pubDate>
		<guid isPermaLink="false">http://www.potencyassay.com/?p=45#comment-9</guid>
		<description>Hello Dan, thank you for this blog. How do you generate the heatmap from your experimental data? Is there a special software?</description>
		<content:encoded><![CDATA[<p>Hello Dan, thank you for this blog. How do you generate the heatmap from your experimental data? Is there a special software?</p>